159 research outputs found

    Improved calibration procedure for laser Doppler perfusion monitors

    No full text
    Commercial laser Doppler perfusion monitors are calibrated using the perfusion value, i.e. the first order moment of the Doppler power spectrum, from a measurement in a standardized microsphere colloidal suspension under Brownian motion. The calibration perfusion value depends on several parameters of the suspension that are difficult to keep constant with adequate accuracy, such as the concentration, temperature and the microsphere size distribution. The calibration procedure itself may therefore introduce significant errors in the measured values. An altered calibration procedure, where the zero order moment is used is described and demonstrated in this paper. Since the above mentioned parameters only affect the frequency content of the Doppler power spectrum and not the total power, the zero order moment will be independent of those parameters. It is shown that the variation in the calibration value, as given by measurements on different scattering liquids with a wide range of scattering properties and temperatures, is only a few percent using the proposed method. For the conventional calibration procedure, this variation corresponds to an error introduced by merely a 1°C variation in the reference liquid temperature. The proposed calibration method also enables absolute level comparisons between measured and simulated Doppler power spectra.Original Publication: Ingemar Fredriksson, M. Larsson, T. Strömberg and F. Salomonsson, Improved calibration procedure for laser Doppler perfusion monitors, 2011. http://dx.doi.org/10.1117/12.871938 Copyright 2011 Society of Photo-Optical Instrumentation Engineers. One print or electronic copy may be made for personal use only. Systematic reproduction and distribution, duplication of any material in this paper for a fee or for commercial purposes, or modification of the content of the paper are prohibited.The publication is also included in the series Progress in Biomedical Optics and Imaging (ISSN 1605-7422) vol. 12 issue 24.</p

    Simulation of reflected light intensity changes during navigation and radio frequency lesioning in the brain

    No full text
    An electrode with adjacent optical fibers for measurements during navigation and radio frequency lesioning in the brain was modeled for Monte Carlo simulations of light transport in brain tissue. Relative reflected light intensity at 780 nm, I780, from this electrode and probes with identical fiber configuration were simulated using the intensity from native white matter as reference. Models were made of homogeneousnative and coagulated gray, thalamus, and white matter as well as blood. Dual layermodels, including models with a layer of cerebrospinal fluid between the fibers andthe brain tissue, were also made. Simulated I780 was 0.16 for gray matter, 0.67 forcoagulate gray matter, 0.36 for thalamus, 0.39 for coagulated thalamus, unity forwhite matter, 0.70 for coagulated white matter and 0.24 for blood. Thalamic matterhas also been found to reflect more light than gray matter and less than white matterin clinical studies. In conclusion the reflected light intensity can be used todifferentiate between gray and white matter during navigation. Furthermore,coagulation of light gray tissue, such as the thalamus, might be difficult to detectusing I780, but coagulation in darker gray tissue should result in a rapid increase of I780.Original Publication: Johannes D. Johansson, Ingemar Fredriksson, Karin Wårdell and Ola Eriksson, Simulation of reflected light intensity changes during navigation and radio frequency lesioning in the brain, Journal of Biomedical Optics, (14), 044040, (2009). http://dx.doi.org/10.1117/1.3210781 Copyright 2009 Society of Photo-Optical Instrumentation Engineers. One print or electronic copy may be made for personal use only. Systematic reproduction and distribution, duplication of any material in this paper for a fee or for commercial purposes, or modification of the content of the paper are prohibited.</p

    Reduced Arteriovenous Shunting Capacity After Local Heating and Redistribution of Baseline Skin Blood Flow in Type 2 Diabetes Assessed With Velocity-Resolved Quantitative Laser Doppler Flowmetry

    No full text
    OBJECTIVE-To compare the microcirculatory velocity distribution in type 2 diabetic patients and nondiabetic control subjects at baseline and after local heating. RESEARCH DESIGN AND METHODS-The skin blood flow response to local heating (44 degrees C for 20 mm) was assessed in 28 diabetic patients and 29 control subjects using a new velocity-resolved quantitative laser Doppler flowmetry technique (qLDF). The qLDF estimates erythrocyte (RBC) perfusion (velocity X concentration), in a physiologically relevant unit (grams RBC per 100 g tissue X millimeters per second) in a fixed output volume, separated into three velocity regions: v less than1 mm/s, v 1-10 mm/s, and v greater than10 mm/s. RESULTS-The increased blood flow occurs in vessels with a velocity greater than1 mm/s. A significantly lower response in qLDF total perfusion was found in diabetic patients than in control subjects after heat provocation because of less high-velocity blood flow (v greater than10 mm/s). The RBC concentration in diabetic patients increased sevenfold for v between 1 and 10 mm/s, and 15-fold for v greater than10 mm/s, whereas no significant increase was found for v less than1 mm/s. The mean velocity increased from 0.94 to 7.3 mm/s in diabetic patients and from 0.83 to 9.7 mm/s in control subjects. CONCLUSIONS-The perfusion increase occurs in larger shunting vessels and not as an increase in capillary flow. Baseline diabetic patient data indicated a redistribution of flow to higher velocity regions, associated with longer duration of diabetes. A lower perfusion was associated with a higher BMI and a lower toe-to-brachial systolic blood pressure ratio.This is an author-created, uncopyedited electronic version of an article accepted for publication in Diabetes. The American Diabetes Association (ADA), publisher of Diabetes, is not responsible for any errors or omissions in this version of the manuscript or any version derived from it by third parties. The definitive publisher-authenticated version is online at http://diabetes.diabetesjournals.org.:Ingemar Fredriksson, Marcus Larsson, Fredrik Nyström, Toste Länne, Carl Johan Östgren and Tomas Strömberg, Reduced Arteriovenous Shunting Capacity After Local Heating and Redistribution of Baseline Skin Blood Flow in Type 2 Diabetes Assessed With Velocity-Resolved Quantitative Laser Doppler Flowmetry, 2010, Diabetes, (59), 7, 1578-1584.http://dx.doi.org/10.2337/db10-0080Copyright: American Diabetes Association Inc / American Diabetes Association; 1999http://www.diabetes.org

    Forced detection Monte Carlo algorithms for accelerated blood vessel image simulations

    No full text
    Two forced detection (FD) variance reduction Monte Carlo algorithms for image simulations of tissue-embedded objects with matched refractive index are presented. The principle of the algorithms is to force a fraction of the photon weight to the detector at each and every scattering event. The fractional weight is given by the probability for the photon to reach the detector without further interactions. Two imaging setups are applied to a tissue model including blood vessels, where the ID algorithms produce identical results as traditional brute force simulations, while being accelerated with two orders of magnitude. Extending the methods to include refraction mismatches is discussed. The principle of forced detection; a part of the photon weight. based on the probability of reaching the detector without further interactions, is forced to the detector at each and every scattering event.This is the pre-peer reviewed version of the following article:Ingemar Fredriksson, Marcus Larsson and Tomas Strömberg, Forced detection Monte Carlo algorithms for accelerated blood vessel image simulations, 2009, JOURNAL OF BIOPHOTONICS, (2), 3, 178-184.which has been published in final form at: http://dx.doi.org/10.1002/jbio.200810048Copyright: Wiley-Blackwel

    Absolute flow velocity components in laser Doppler flowmetry

    No full text
    A method to separate a Doppler power spectrum into a number of flow velocity components, measured in absolute units (mm/s), is presented. A Monte Carlo software was developed to track each individual Doppler shift, to determine the probability, p(n), for a photon to undergo n Doppler shifts. Given this shift distribution, a mathematical relationship was developed and used to calculate a Doppler power spectrum originating from a certain combination of velocity components. The non linear Levenberg-Marquardt optimization method could thus be used to fit the calculated and measured Doppler power spectra, giving the true set of velocity components in the measured sample. The method was evaluated using a multi tube flow phantom perfused with either polystyrene microspheres or undiluted/diluted human blood (hct = 0.45). It estimated the velocity components in the flow phantom well, during both low and high concentrations of moving scatterers (microspheres or blood). Thus, further development of the method could prove to be a valuable clinical tool to differentiate capillary blood flow.Ingemar Fredriksson, Marcus Larsson and Tomas Strömberg, Absolute flow velocity components in laser Doppler flowmetry, 2006, Proceedings of SPIE -- Volume 6094 Optical Diagnostics and Sensing VI. http://dx.doi.org/10.1117/12.659206. Copyright 2006 Society of Photo-Optical Instrumentation Engineers. This paper was published in Proceedings of SPIE -- Volume 6094 Optical Diagnostics and Sensing VI and is made available as an electronic reprint with permission of SPIE. One print or electronic copy may be made for personal use only. Systematic or multiple reproduction, distribution to multiple locations via electronic or other means, duplication of any material in this paper for a fee or for commercial purposes, or modification of the content of the paper are prohibited

    Optical microcirculatory skin model: Assessed by Monte Carlo simulations paired with in vivo laser Doppler flowmetry

    No full text
    An optical microvascular skin model, valid at 780 nm, was developed. The model consisted of six layers with individual optical properties, and variable thicknesses and blood concentrations at three different blood flow velocities. Monte Carlo simulations were used to evaluate the impact of various model parameters on the traditional Laser Doppler flowmetry (LDF) measures. A set of reference Doppler power spectra was generated by simulating 7,000 configurations, varying the thickness and blood concentrations. Simulated spectra, at two different source detector separations, were compared with in vivo recorded spectra, using a non-linear search algorithm for minimizing the deviation between simulated and measured spectra. The model was validated by inspecting the thickness and blood concentrations which generated the best fit. These four parameters followed a priori expectations for the measurement situations, and the simulated spectra agreed well with the measured spectra for both detector separations. Average estimated dermal blood concentration was 0.08% at rest and 0.63% during heat provocation (44°C) on the volar side of the forearm, and 1.2% at rest on the finger pulp. The model is crucial for developing a technique for velocity-resolved absolute LDF measurements with known sampling volume, and can also be useful for other bio-optical modalities.Ingemar Fredriksson, Marcus Larsson and Tomas Strömberg, Optical microcirculatory skin model: Assessed by Monte Carlo simulations paired with in vivo laser Doppler flowmetry, 2008, Journal of Biomedical Optics, (13), 1, 14015. http://dx.doi.org/10.1117/1.2854691. Copyright 2008 Society of Photo-Optical Instrumentation Engineers. This paper was published in Journal of Biomedical Optics and is made available as an electronic reprint with permission of SPIE. One print or electronic copy may be made for personal use only. Systematic or multiple reproduction, distribution to multiple locations via electronic or other means, duplication of any material in this paper for a fee or for commercial purposes, or modification of the content of the paper are prohibited

    Measurement depth and volume in laser Doppler flowmetry

    No full text
    A new method for estimating the measurement depth and volume in laser Doppler flowmetry (LDF) is presented. The method is based on Monte Carlo simulations of light propagation in tissue. The contribution from each individual Doppler shift is calculated and thereby multiple Doppler shifts are handled correctly. Different LDF setups for both probe based (0.0, 0.25, 0.5, and 1.2 mm source-detector separation) and imaging systems (0.5 and 2.0 mm beam diameter) are considered, at the wavelengths 543 nm, 633 nm, and 780 nm. Non-linear speckle pattern effects are accounted for in the imaging system setups. The effects of tissue optical properties, blood concentration, and blood oxygen saturation are evaluated using both homogeneous tissue models and a layered skin model. The results show that the effect on the measurement depth of changing tissue properties is comparable to the effect of changing the system setup, e.g. source-detector separation and wavelength. Skin pigmentation was found to have a negligible effect on the measurement depth. Examples of measurement depths are (values are given for a probe based system with 0.25 mm source-detector separation and an imaging system with a 0.5 mm beam diameter, respectively, both operating at 780 nm): muscle - 0.55/0.79 mm; liver - 0.40/0.53 mm; gray matter - 0.48/0.68 mm; white matter - 0.20/0.20 mm; index finger pulp - 0.41/0.53 mm; forearm skin - 0.53/0.56 mm; heat provoked forearm skin - 0.66/0.67 mm.Original Publication:Ingemar Fredriksson, Marcus Larsson and Tomas Strömberg, Measurement depth and volume in laser Doppler flowmetry, 2009, Microvascular Research, (78), 1, 4-13.http://dx.doi.org/10.1016/j.mvr.2009.02.008Copyright: Elsevier Science B.V., Amsterdamhttp://www.elsevier.com

    Corresponding author:

    No full text
    N.B.: When citing this work, cite the original article. This is an author-created, uncopyedited electronic version of an article accepted for publication in Diabetes. The American Diabetes Association (ADA), publisher of Diabetes, is not responsible for any errors or omissions in this version of the manuscript or any version derived from it by third parties. The definitive publisher-authenticated version is online a

    Quantitative Laser Doppler Flowmetry

    No full text
    Laser Doppler flowmetry (LDF) is virtually the only non-invasive technique, except for other laser speckle based techniques, that enables estimation of the microcirculatory blood flow. The technique was introduced into the field of biomedical engineering in the 1970s, and a rapid evolvement followed during the 1980s with fiber based systems and improved signal analysis. The first imaging systems were presented in the beginning of the 1990s. Conventional LDF, although unique in many aspects and elegant as a method, is accompanied by a number of limitations that may have reduced the clinical impact of the technique. The analysis model published by Bonner and Nossal in 1981, which is the basis for conventional LDF, is limited to measurements given in arbitrary and relative units, unknown and non-constant measurement volume, non-linearities at increased blood tissue fractions, and a relative average velocity estimate. In this thesis a new LDF analysis method, quantitative LDF, is presented. The method is based on recent models for light-tissue interaction, comprising the current knowledge of tissue structure and optical properties, making it fundamentally different from the Bonner and Nossal model. Furthermore and most importantly, the method eliminates or highly reduces the limitations mentioned above. Central to quantitative LDF is Monte Carlo (MC) simulations of light transport in tissue models, including multiple Doppler shifts by red blood cells (RBC). MC was used in the first proof-of-concept study where the principles of the quantitative LDF were tested using plastic flow phantoms. An optically and physiologically relevant skin model suitable for MC was then developed. MC simulations of that model as well as of homogeneous tissue relevant models were used to evaluate the measurement depth and volume of conventional LDF systems. Moreover, a variance reduction technique enabling the reduction of simulation times in orders of magnitudes for imaging based MC setups was presented. The principle of the quantitative LDF method is to solve the reverse engineering problem of matching measured and calculated Doppler power spectra at two different source-detector separations. The forward problem of calculating the Doppler power spectra from a model is solved by mixing optical Doppler spectra, based on the scattering phase functions and the velocity distribution of the RBC, from various layers in the model and for various amounts of Doppler shifts. The Doppler shift distribution is calculated based on the scattering coefficient of the RBC:s and the path length distribution of the photons in the model, where the latter is given from a few basal MC simulations. When a proper spectral matching is found, via iterative model parameters updates, the absolute measurement data are given directly from the model. The concentration is given in g RBC/100 g tissue, velocities in mm/s, and perfusion in g RBC/100 g tissue × mm/s. The RBC perfusion is separated into three velocity regions, below 1 mm/s, between 1 and 10 mm/s, and above 10 mm/s. Furthermore, the measures are given for a constant output volume of a 3 mm3 half sphere, i.e. within 1.13 mm from the light emitting fiber of the measurement probe. The quantitative LDF method was used in a study on microcirculatory changes in type 2 diabetes. It was concluded that the perfusion response to a local increase in skin temperature, a response that is reduced in diabetes, is a process involving only intermediate and high flow velocities and thus relatively large vessels in the microcirculation. The increased flow in higher velocities was expected, but could not previously be demonstrated with conventional LDF. The lack of increase in low velocity flow indicates a normal metabolic demand during heating. Furthermore, a correlation between the perfusion at low and intermediate flow velocities and diabetes duration was found. Interestingly, these correlations were opposites (negative for the low velocity region and positive for the mediate velocity region). This finding is well in line with the increased shunt flow and reduced nutritive capillary flow that has previously been observed in diabetes
    corecore