    Water Triple Point Comparison: Plateau averaging or Peak Value?

    No full text
    With a certain regularity, national metrology institutes conduct comparisons of water triple-point (WTP) cells. The WTP is the most important fixed point for the International Temperature Scale of 1990 (ITS-90). In such comparisons, it is common practice to simply average all the single measured temperature points obtained on a single ice mantle. This practice is quite reasonable whenever the measurements show no time dependence in the results. Ever since the first Supplementary Information for the International Temperature Scale of 1990, published by the Bureau International des Poids et Mesures in Sèvres, it was strongly suggested to wait at least 1 day before taking measurements (now up to 10 days), in order for a newly created ice mantle to stabilize. This stabilization is accompanied by a change in temperature with time. A recent improvement in the sensitivity of resistance measurement enabled the Istituto Nazionale di Ricerca Metrologica to detect more clearly the (possible) change in temperature with time of the WTP on a single ice mantle, as for old borosilicate cells. A limited investigation was performed where the temperature of two cells was monitored day-by-day, from the moment of mantle creation, where it was found that with (old) borosilicate cells it may be counterproductive to wait the usual week before starting measurements. The results are presented and discussed, and it is suggested to adapt the standard procedure for comparisons of WTP cells allowing for a different data treatment with (old) borosilicate cells, because taking the temperature dependence into account will surely reduce the reported differences between cells

    Production of a new Tin Cell at INRIM

    No full text
    Over the years some of the INRIM metal fixed-point cells, ≥20 years old, have shown a variable deterioration consisting in the leaking of some of the metal through the walls of the crucible. The variation in the deterioration is assumed to be connected with the frequency with which the cells have been used. New cells are therefore being produced at INRiM as substitutes for the older cells. A description of the new cells and their preparation is given, along with a preliminary comparison between the new cell and the older cell(s)

    Some Curious Results with a Gallium Fixed Point Cell

    No full text
    In 2009, most of the gallium fixed-point cells in use in different INRIM laboratories were compared with Italy’s national standard. The comparison has uncovered problems with one of the commercial devices, realizing a temperature about 0.7mK too low which initially was even changing linearly with time. An additional series of measurements was undertaken to find out the reason for this behavior, but not being allowed to open the cell, only a suspicion on the possible cause has remained. A way is suggested that might give users an indication of such misbehavior of their cell. The results underline the importance for those NMIs with only a single cell, for any fixed point, to undertake regular comparisons with another cell as a check on its behavior
    corecore